Stability band width ( 5t-b ), Zero tracking ( trc ), Display refresh rate ( 5pd ) – A&D Weighing HR-202i User Manual
Page 32: Decimal point ( pnt ), Auto display-on ( p-on )
30
10.3.
Description of the Class "Environment, Display"
Condition ( Cond )
Cond 0
This parameter is for sensitive response to the fluctuation of a mass value. Used
for powder target weighing, weighing a very light sample or when quick response
weighing is required. After setting, the balance displays
FAST
.
Cond 2
This parameter is for stable weighing with slow response. Used to prevent a mass
value from drifting due to vibration or drafts. After setting, the balance displays
SLOW
.
Stability band width ( 5t-b )
This item controls the width to regard a mass value as a stable value. When the fluctuation
per second is less than the parameter, the balance displays the stabilization indicator and
outputs the data. The parameter influences the "Auto print mode"
5t-b
0 This parameter is used for sensitive response of the stabilization indicator. Used
for exact weighing.
5t-b
2 This parameter ignores slight fluctuations of a mass value. Used to prevent a
mass value from drifting due to vibration or drafts.
Zero tracking ( trc )
This function tracks zero point drift caused by changes in the environment and stabilizes the
zero point. When the weighing data is only a few digits, turn the function off for accurate
weighing.
trc 0
The tracking function is not used. Used for weighing a very light sample.
trc 1
The normal tracking function is used.
trc 2
The strong tracking function is used.
trc 3
The very strong tracking function is used. Used for stable zero display.
Display refresh rate ( 5pd )
The period to refresh the display. This parameter influences "Baud rate", "Data output
pause" and "Stream mode".
Decimal point ( pnt )
The decimal point format can be selected.
Auto display-ON ( p-on )
When the AC adapter is plugged in, the display is automatically turned on without the
ON:OFF
key operation, to display the weighing mode. Used when the balance is built into
an automated system. one hour warm up is necessary for accurate weighing.