beautypg.com

Function switch and initialization, Permit or inhibit – A&D Weighing HR-202i User Manual

Page 25

background image

23

9. Function

Switch and Initialization

9.1.

Permit or Inhibit

The balance stores parameters that must not be changed unintentionally (Example:

Calibration data for accurate weighing, Data for adapting to the operating environment,
Control data for the RS-232C interface). There are two switches for the purpose of
protecting parameters. Each switch can select either "permit" or "inhibit". "Inhibit" protects
parameters against unintentional operations.


1 Press

the

ON:OFF

key to turn off the display.


2 While pressing and holding the

PRINT

key and the

RANGE

key,

press the

ON:OFF

key to display

p5

.


3 Press

the

PRINT

key. Then the balance displays the function switches.


4 Specify the switches using the following keys.

RANGE

key ...........The key to select blinking digit.

RE-ZERO

key.........The key to change the parameter for the selected switch.

0 To inhibit changes. (Can not be used.)

1 To permit changes. (Can be used.)

PRINT

key .............The key to store the new parameter and return to the weighing mode.

CAL

key.................The key to cancel the current operation and return to the weighing

mode.

















Function table

0 To inhibit changes to the function table.

1 To permit changes to the function table.

Calibration using the external weight

0 To inhibit calibration using the external weight.

1 To permit calibration using the external weight.