Function switch and initialization, Permit or inhibit – A&D Weighing HR-202i User Manual
Page 25
23
9. Function
Switch and Initialization
9.1.
Permit or Inhibit
The balance stores parameters that must not be changed unintentionally (Example:
Calibration data for accurate weighing, Data for adapting to the operating environment,
Control data for the RS-232C interface). There are two switches for the purpose of
protecting parameters. Each switch can select either "permit" or "inhibit". "Inhibit" protects
parameters against unintentional operations.
1 Press
the
ON:OFF
key to turn off the display.
2 While pressing and holding the
key and the
RANGE
key,
press the
ON:OFF
key to display
p5
.
3 Press
the
key. Then the balance displays the function switches.
4 Specify the switches using the following keys.
RANGE
key ...........The key to select blinking digit.
RE-ZERO
key.........The key to change the parameter for the selected switch.
0 To inhibit changes. (Can not be used.)
1 To permit changes. (Can be used.)
key .............The key to store the new parameter and return to the weighing mode.
CAL
key.................The key to cancel the current operation and return to the weighing
mode.
Function table
0 To inhibit changes to the function table.
1 To permit changes to the function table.
Calibration using the external weight
0 To inhibit calibration using the external weight.
1 To permit calibration using the external weight.