6-10

Chapter 6. Calibration

Note

Changing the values in the linearization
table incorrectly can cause the Weigh
II to display incorrect sensor data. Do
not change the values in the lineariza-
tion table unless one of the above
problems has been noted.

An example is provided below of a situation
where use of the linearization table may be
required:

Example: When your vessel is empty, the
Weigh II correctly shows approximately 0 lbs.
You start putting truckloads of material into
the vessel, and notice the following pattern:

Truck

Total Actual

Total Indicated

Load

Load

0

0

1

5,000

6,000

2

10,000

11,500

3

15,000

16,000

4

20,000

20,000

After the fourth truckload, your indicated load
from the Weigh II is reasonably accurate.
You notice this same pattern and magnitude
of errors every time you start with an
empty vessel.

The Weigh II’s linearization algorithm uses a
five-point piece-wise linearization method with
linear interpolation between points. Figure 6-4
illustrates the linearization operation. Table
6-1 is the default linearization table, consist-
ing of five raw digital count values as inputs
and five corrected digital count values as
outputs. Note that for the default condition,
the raw and corrected values are identical, so
the linearization table will have no effect on
the Weigh II, even if linearization is On.

Raw Value

Corrected Value

1048576

1048576

1112576

1112576

1176576

1176576

1240576

1240576

1304576

1304576

Table 6-1. Default Linearization Table

Consult with Kistler-Morse to determine the
values to input for linearization.

Follow this procedure to enable, disable, and
set up the linearization table.

1. If the Weigh II is in Auto Mode (Auto LED

illuminated), press the Auto/Man Key to
put the system in Manual Mode. The Auto
LED will turn off.

2. Use the Up Arrow or Down Arrow Key to

scroll to the desired vessel.

3. Press the Menu Key to display the Main

Menu. The display shows:

4. Press the F3 Key to access the Cal Menu.

The display shows:

5. Press the Menu Key again to display the

second page of the menu. The display
shows:

6. Press the F1 Key to access the Linear

Menu. The display shows:

An asterisk indicates the current selection
for On or Off.

7. Press the F1 Key to turn the linearization

table on or the F2 Key to turn the table off.
The display flashes a message acknowl-
edging your selection and
returns to:

An asterisk indicates the new selection for
On or Off.

If you want to change the linearization
table, continue with Step 8. If not, press
the Esc Key to scroll up the menu tree or
press the Auto/Man Key to return the
display to vessel monitoring.

CALIBRATION MENU

Auto Manual

F1 F2 F3

CALIBRATION MENU

Linear

F1 F2 F3

WII MAIN MENU

Disp I/O Cal

F1 F2 F3

LINEARITY MENU

On Off* Set

F1 F2 F3

LINEARITY MENU

On* Off Set

F1 F2 F3