Explanation of features, Active tuning, Ambient pressure compensation – KEYENCE AP-C40W(P) Series User Manual

Page 4

background image

4

10. Explanation of Features

Focus mode

Focus mode increases all display resolutions by a factor of 10.

Normal mode

Focus mode

The following focus center pressures can be selected for the models
noted in the table below.

(Unit: kPa)

Head

41
43
44
48
43
48
41
44
41
44
41
43
44
48
41
43
44
48

-20,0/-30,0/-40,0/-50,0/-60,0/-70,0/-80,0

200/300/400/500/600/700/800

80,0/60,0/40,0/20,0/0,0/-20,0/-40,0/-60,0/-80,0
-80,0/-60,0/-40,0/-20,0/0,0/20,0/40,0/60,0/80,0

2,00/3,00/4,00/5,00/6,00/7,00/8,00

-800/-600/-400/-200/0/200/400/600/800

-100/-200/-300/-400/-500/-600/-700

600/500/400/200/0/-200/-400/-500/-600

0,0/-5,0/-10,0/-15,0/-20,0/-25,0/-30,0

20,0/15,0/10,0/5,0/0,0/-5,0/-10,0/-15,0/-20,0

-1,00/-3,00/-5,00/-7,00/-9,00/-11,00/-13,00

10,0/30,0/50,0/70,0/90,0/110,0/130,0

12,00/9,00/6,00/3,00/0,00/-3,00/-6,00/-9,00/-12,00
-12,00/-9,00/-6,00/-3,00/0,00/3,00/6,00/9,00/12,00

-200/-300/-400/-500/-600/-700/-800

2,00/3,00/4,00/5,00/6,00/7,00/8,00

800/600/400/200/0/-200/-400/-600/-800
-800/-600/-400/-200/0/200/400/600/800

Selected units

bar

psi

inH

mmH

GF

Pa

kPa

mbar

mbar

kgf/cm

2

gf/cm

2

mmHg

psi

inHg

bar

Unit

Center Pressure

* The current value is displayed as ±20% of the F.S. range with the

focus center pressure as the reference value (0).

Zero shift

Zero shift input forces the pressure at that time to be defined as zero.
This feature is useful in applications that require the detection of a cer-
tain amount of pressure fluctuation without being influenced by changes
in the original pressure.

0

-0.1MPa
O.K.

Zero shift input

Zero shift input

-0.5MPa
N.G.

1MPa

(Example: leak testing)
Zero shift input received when a container is filled with air will allow the
amount of leak after a certain time to be displayed as negative pressure.
This approach eliminates the influence of small variations in the final fill
pressure of the container.

The pressure value (reference value) when zero shift input is received
can be verified from the current value display by pressing the button
to switch to the reference value display.
(

)

Press the button once more to return to the current value display.

Analog output

A voltage corresponding to the pressure value is output.
* Ambient pressure compensation is also applied to analog output.

Head

1 V

to

5 V

AP-41/41M

0

to -101.3 kPa

AP-43

0

to

+1 MPa

AP-44/48

-101.3 to +101.3 kPa

* The resolution of analog voltage output will not be enhanced during

focus mode operation.

Peak/bottom hold display

The maximum (minimum) values are displayed continuously after power to
the sensor is turned on.
The active hold display indicated below is used when the A-1 mode is selected.

How to display hold values/

From current value/reference value display

+

/ Activates peak hold display.

+

/Activates bottom hold display.

The minimum value during zero shift input is displayed for the bottom
hold during A-2 mode operation.

Resetting the peak/bottom hold values/

Pressing

for 3 seconds or more while the peak and bottom hold

values are being displayed will reset the values.
*The values will also be reset when power to the sensor is turned off,

and when the device's detection mode or operating mode is changed.

Power-save

The value display will be turned off during power-save operation to re-
duce the amount of power consumed.

Use of any keys during eco mode operation will revert the sensor to
its normal display. The sensor will return to the eco display when
there is no key input for a period of 3 minutes.

Switching the display color

Corresponds to either output 1 or output 2 based on which setting dis-
play has been selected.
* rGr can only be set when either the F-1 or F-2 mode is selected.

Values that fall between the P1 and P2 settings will be displayed in
green, while values outside that range will be displayed in red.

11. Active Tuning

Active 2 point tuning

Active 1 point tuning

Suction

Vacuum burst

P1

P2

T1

Automatic

configuration

_

_

Automatic

configuration

Manual

configuration

_

Application

Type

P2

P1

Time

Zero shift input

T1

V

acuum

Perform these steps first

Select the setting display as indicated below before performing the
tuning procedures.
Active 2 point tuning: Select either P1 or T1 for the setting display.
Active 1 point tuning: Select P2 from the settings display.

Active 2 point tuning / a tuning method suited for auto-
matically configuring suction pick-up detection pressures

The zero shift timer setting is manually configured, and the sensor au-
tomatically selects the best pressure setting (P1) for this value.

Step 1)

Step 2)

Device operation

with no target

object present

Operation

for several

cycles

Operation

for several

cycles

Device operation

with target

object present

Sensitivity

difference

display

Sampling

Sampling

Start

Start

Finish

Finish

Active 1 point tuning / tuning optimized for vacuum burst
detection

The sensor automatically selects the best pressure setting (P2) for
vacuum burst detection.

Target object

present

Setting

display

Sampling

Start

Finish

12. Ambient Pressure Compensation

Open the applied pressure to the surrounding air so that it equals the
ambient pressure and press the

button for 3 seconds or more while

the current value or reference value is being displayed.
The ambient pressure point will be corrected.

During A-1 mode operation (active hold display)

The hold values are reset each time the pressure value exceeds (or falls
below) the setting, and the peak hold (bottom hold) will then vary from
this point.

Peak hold value:

The peak hold value will be reset once a value is
encountered that exceeds setting P1, and new peak
values will be held from that point.

Bottom hold value: The bottom hold value will be reset once a value is

encountered that falls below setting P2, and new
bottom values will be held from that point.