Parameters of commands and responses, Coriolis digital flow sensor fd-s series – KEYENCE DL-RS1A/FD-S User Manual
Page 12
10
Parameters of Commands and Responses
This section describes the parameters used with various commands and responses.
Coriolis Digital Flow Sensor FD-S Series
■ Communication commands
There are two types of communication commands: read commands and write commands.
■ ID numbers
This parameter is used with communication commands "SR" and "SW". Specify the ID number of the
target sensor amplifier using two digits (ASCII characters).
Example: Sensor amplifier of ID number 00 = 00, sensor amplifier of ID number 02 = 02
Communication
command
Attribute
Description
SR
Read
Reads the data for the specified data number from the sensor amplifier with the
specified ID number.
M0
Reads all of the instantaneous flow rate (current value) data that DL-RS1A
periodically retrieves from all sensor amplifiers.
MS
Reads the control output state and instantaneous flow rate (current value) data
that DL-RS1A periodically retrieves from all sensor amplifiers.
DR
Reads the control output state and instantaneous flow rate (current value) data
that DL-RS1A periodically retrieves from all sensor amplifiers when a DRQ input
is received.
SW
Write
Writes setting data for the specified data number into the sensor amplifier with
the specified ID number.
AW
Writes setting data for the specified data number into all sensor amplifiers.
Parameters of Commands and Responses
■ Data numbers
Specify the data number using three digits (ASCII characters).
●
Read-only data
The following table lists the types of data that can only be read from FD-S Series sensor amplifiers.
Writing read-only data results in a communication error (error number: 22).
*1
In the Data format column, "*" signifies a number from "0 to 9".
*2
R: Indicates that the data type can only be read from the sensor amplifiers.
*3
Regardless of the display mode setting specified on the sensor amplifier, this is the data during
the “rESo” mode.
*4
Regardless of the integration display unit specified on the sensor amplifiers, the data is read
with the minimum display unit.
When the integrated flow quantity exceeds the indicated maximum value, the maximum value is
returned as the data.
Data
number
Data name
Data format
*1
Data range
0.2L
1L, 2L
8L, 20L
0.2L
1L
2L
8L
20L
000
Instantaneous
flow rate
±***.*
±****
±**.**
-400.0 to
+400.0
*2
-2000 to
+2000
-4000 to
+4000
-16.00 to
+16.00
-40.00 to
+40.00
001
Integrated flow
quantity
(Original Method)
±********.*
±*********
±*******.**
-21474836.4 to
+21474836.4
*4
-214748364 to
+214748364
-2147483.64 to
+2147483.64
002
Instantaneous
flow rate
(peak hold value)
±***.*
±****
±**.**
-400.0 to
+400.0
*3
-2000 to
+2000
-4000 to
+4000
-16.00 to
+16.00
-40.00 to
+40.00
003
Instantaneous
flow rate
(bottom hold
value)
-400.0 to
+400.0
*3
-2000 to
+2000
-4000 to
+4000
-16.00 to
+16.00
-40.00 to
+40.00
004
Instantaneous
density
**.**
00.00 to 99.99
005
Output terminal
state
*
Bit0: output 1 (ON at 1)
Bit1: output 2 (ON at 1)
Bit2: output 3 (ON at 1)
006
Input terminal
state
0: OFF, 1: ON
007
Bank switching
state
0: bank A; 1: bank B
008
Error state
****
0 to 2604
*5
009
Connected
sensor head
(quality)
*
0: SUS
1: PFA
1: PFA
0: SUS
1: PFA
1: PFA
0: SUS
010
Connected
sensor head
(capacity)
0: 0.2L
1: 2L(Foc)
3: L(Foc)
3ÅF1L
1: 2L
2: 20L(Foc)
4: 8L(Foc)
4: 8L
2: 20L
011
Integrated flow
quantity
(Corresponding
to the A-Ver)
±********.**
±*********.*
±*******.***
-21474836.48 to
+21474836.47
-214748364.8 to
+214748364.7
-2147483.648 to
+2147483.647
015
Temperature
(current value)
±***.*
000.0 to 099.9
*6
016
Temperature
(peak hold value)
017
Temperature
(bottom hold value)
019
Shift system
processing results
*
0: normal termination; 1: abnormal termination
Point