Holtgreven Ohaus Defender Series Bench Scales User Manual
Page 34
EN-32
5000 Series Indicators
Reset No, Yes
Stable Only Off, On
Auto Print Off,
On Stable (-> Load, Load and Zero),
Interval (-> 0…3600), Continuous,
On Accept
Print Content Result (-> Off, On, Numeric only),
Gross (-> Off, On),
Net (-> Off, On),
Tare (-> Off, On),
Header ( ->Off, On),
User ID (-> Off, On),
Project ID (-> Off, On),
Scale ID (-> Off, On),
Difference (-> Off, On),
Date and Time (-> Off, On),
Information (-> Off, On),
Application Mode ( Off, On),
Name (-> Off, On),
Layout
Format (-> Multiple, Single),
Feed (-> Line feed, 4 Line feed, Form
feed)
List No, Yes
End Print1 Exit PRINT1 menu
(End Print2) Exit PRINT2 menu
3.9.1 Reset
Set the Print menu to factory defaults.
NO
= not reset.
YES
= reset.
3.9.2 Print Stable Data Only
Set the print criteria.
OFF = values are printed immediately.
ON = values are only printed when the stability criteria are met.
3.9.3 Auto Print
Set the automatic printing functionality.
OFF
= disabled.
ON.StAb = printing occurs each time the stability criteria are met.
INtEr = printing occurs at the defined interval.
CONt
= printing occurs continuously.
ACCEPt = printing occurs each time the display is within the Checkweigh accept range and
stability criteria are met.
3.9 Print1 and Print2 Menus
Enter this menu to define printing parameters. Default settings are bold.
NOTE: The Print2 menu is only displayed if a second interface
(RS232 or RS422/RS485) is installed.
NOTE: If the Legal for Trade menu item is set to ON, the following
settings are not reset: Stable