beautypg.com

Holtgreven Ohaus Defender Series Bench Scales User Manual

Page 20

background image

EN-18

5000 Series Indicators

3.3.5 Geographical Adjustment Factor

Refer to Table 3-2 and set the GEO factor that corresponds to your location.
00 to 31

NOTE: Only an authorized manufacturer’s representative or certified verification personnel may make these
changes. Changing the geographical setting alters the calibration values.

Advance to the next menu.

3.3.6 End Calibration

3.3.4 Calibration Test

Calibration test is used to compare a known calibration weight against the stored span calibration data.

NOTE: Calibration Test is always available (even when LFT is set to ON).

When tESt is displayed, press the Yes button to initiate Calibration Test.

The display flashes 0. With no weight on the scale, press the Yes button to record the current zero point.

The display shows --t-- while the zero point is recorded.

The display flashes the span calibration weight using the value from the last calibration. The example shows
test weight of 30 kg.

Place the specified test weight on the scale and press the Yes button.
The display shows --t-- while the data is processed.

The display flashes the actual difference between the calibration data and the test weight.

The example shows a 0.010 kg difference. The result of the Calibration Test is printed.

After 5 seconds, Calibration Test ends, the scale returns to the active weighing mode and displays the current
weight.

Е