Hardy HI 1769-FC User Manual
Page 35

30
●
●
●
●
●
Chapter 4
Command
Hex
Description
RATE_CAL_5POI
NT_CMD
0x60
Runs the 5-point rate calibration at whatever points have
previously been defined.
RATE_CAL_2POI
NT_CMD
0x61
Runs the 2-point rate calibration at whatever points have
previously been defined.
RATE_CAL_AD
VANCED
0x62
This command fills in the 7 calibration points with values
extrapolated from 2 entered points, 2 and 4 of a 5-point
calibration. The 2 percentages are entered at offsets 5 and
6 as in the
RATE_CAL_2_POINTS_DEFINED command,
and
the 2 rates are entered at offsets 7 and 8. The rates have
“rate decimal places”, as set in the FORMAT word at
offset 1.
CALLOWCMD
0x64
Sets the “calLowCount” parameter to the current A/D
average counts when doing a hard calibration. Calls an
Integrated Technician function during low calibration.
Error Return Values:
STATUSWORD - There was a conversion error or A/D
conversion failure.
CALHIGHCMD
0x65
Sets the Span Weight parameter to the current A/D
average counts when doing Hard CAL.
Error Return Values:
STATUSWORD - There was a conversion error or A/D
conversion failure.
HARDCALFAILCOUNTS - There are less than 1000
counts between the zero and the span weights.
C2CALCMD
0x66
Performs a C2 Calibration. Uses CalLowWeight as the
reference point.
Error Return Values:
STATUSWORD - There was a conversion error or A/D
conversion failure.
C2FAILNODEVS – Detected no C2 load cells.
C2FAILCAPEQ - Detected two load cells with different
capacities.
WRITEPARAM
0x67
Writes the parameters. See Parameter Table below.
Return Data:
COMMAND STATUS - The status is 0 if the command
succeeded. If a range error was detected. See Status Table
below.
READPARAM
0x69
Reads the parameters. See Parameter Table below.
Return Data:
COMMAND STATUS See Status Table below.
PARAMETERS
SETDEFAULTPA
RAMS
0x94
Loads the default settings into RAM, and non-volatile
RAM. Error Return Values: None