Series TT05 Digital Torque Testers
User’s Guide
24
14.4 Factory Settings
Parameter Setting
Set points
Upper
Disabled (defaults to 80% of full scale, clockwise, when enabled)
Lower
Disabled (defaults to 40% of full scale, clockwise, when enabled)
Filters
Current
8
Displayed
1024
DATA Key Functions
RS-232/USB Output
Enabled
Mitutoyo Output
Disabled
Memory Storage
Enabled
Backlight Auto
Minutes
1
Serial/USB
RS-232 Output Selected
Disabled
USB Output Selected
Enabled
Baud Rate
9,600
Data Format
Numeric + units
Mitutoyo BCD Output
Disabled
Break Detection
Automatic Output
RS-232/USB Output
Disabled
Mitutoyo Output
Disabled
Automatic Storage
Disabled
Automatic Zero
Disabled
Break Detection Settings
Trigger Threshold
10%
% Drop
50%
Auto Zero Delay
5 sec.
1
st
/2
nd
Peak
Disabled
Peak Settings
Threshold 1
10%
% Drop 1
50%
Threshold 2
10%
% Drop 2
50%
Auto Store Peak 1
Disabled
Auto Store Peak 2
Disabled
Auto Zero
Disabled
Automatic Shutoff
Enabled
Minutes
5
Beeps
Keys
Enabled
Alerts
Enabled
Set Points
Momentary
LCD Contrast
10
Initial Mode
Real Time
Units lbFin
Passwords
All passwords disabled