12.2 Settings
The display will appear as follows:
Thresholds and percentage drops are illustrated below:
11.3 Auto Output Settings
Select the output type. Select RS-232/USB and/or Mitutoyo outputs, and select First and/or Second
peaks. The display will appear as follows:
Threshold 1
Sets the percentage of full scale at which the first/second peak detection
feature becomes active. This threshold is provided to ignore peaks that can
occur during sample handling. Available settings: 1-90%, in 1% increments
between 1-5%, and in 5% increments between 5-90%.
Percentage Drop 1 Identifies the first peak via detection of a specified percentage drop from
peak.
Threshold 2
Same as with Threshold 1, but refers to a percentage of full scale beyond
the first peak. For example, for a 50 lbFin capacity tester, if the first peak is
20 lbFin, and Threshold 2 is set to 15%, the threshold is equal to 27.5 lbFin.
Percentage Drop 2 Same as Percentage Drop 1, for the second peak.
Auto Zero Delay
Sets the time delay before the primary and peak readings are zeroed.
Available settings: 1-60 sec., in 1 sec. increments between 1-5 sec., and in
5 sec. increments between 5-60 sec.
PEAK SETTINGS
Thresh. 1:
10 %
% Drop 1: 50 %
Thresh. 2: 10 %
% Drop 2: 50 %
Auto Zero Delay
5 sec.
AUTO OUTPUT
SETTINGS
RS232/USB Output
Mitutoyo Output
First Peak
Second Peak