Model ESM301 / ESM301L Version 2 Test Stand

User’s Guide

16

UNITS:

in/min

KEYS:

maintained

PASSWORD: 0000 (off)


Default setting: off
Available settings: off, on



Label Description

+ or –

Cycles through the available settings

ENTR

Returns to the Test Parameter Setup menu

ESC

Exits the parameter without saving changes

4.16 Password (NEW PWORD)

If desired, a password can be set to prevent unwanted changes to test parameters. The password can be
set to any number between 0000 and 9999. The default setting of 0000 indicates that the password is
disabled, and that the user can freely enter the Test Parameter Setup menu. If misplaced or forgotten, the
password may be reset through a code activation process. Details on this may be found in Section 9.

Default setting: 0000 (off)
Available settings: 0000 – 9999



Label Description

+ or –

Cycles through the available settings

ENTR

Returns to the Test Parameter Setup menu

ESC

Exits the parameter without saving changes

5 OPERATING MODES

5.1 Mode Overview

The ESM301 can be operated in several modes, including combinations of these modes:

1. Basic Mode

Manual control of crosshead motion.


2. Auto Return Mode

Crosshead moves to a limit switch or force set point, travel distance, preload, or sample
break (referred to as soft limits), whichever occurs first. Then, it reverses and moves at
maximum speed to the other limit, whichever occurs first.


3. Cycling Mode

Crosshead cycles between limits at the selected speed(s), and pauses at each limit or set
point for a selected period of time.


4. Preload Mode

Crosshead moves until the preload has been reached, and performs the action programmed
in the setting. An auto-return, cycle/dwell time, or break detection sequence may follow.

D E F A U L T ? : N O

E S C

− + E N T R

N E W P WO R D 0 0 0 0

E S C

− + E N T R