Brookfield Bath, Model TC-602 User Manual
Page 17
16
Programming Examples
Example A
Initial Bath Temperature = 25°C
Program Priority = Temperature
Desired Profile: Cool bath temperature to 20°C and hold it there of 25 minutes.
Increase bath temperature to 30°C and hold it there for 15 minutes.
Increase bath temperature to 50°C over a 45 minute period.
Decrease bath temperature to 5°C and hold.
This example requires a 6-step program:
Program
Step
Step Set
Point
Step
Duration
Controller Operation
1
20°C
5 minutes
Controller cools fluid to 20°C as fast as possible. Until
20°C temperature set point is achieved, “Starting”
appears on the display. When fluid temperature reaches
20°C, “Running” appears on the display.
Fluid temperature maintained at 20°C for 25 minutes.
2
20°C
1 second
Controller heats fluid as fast as possible until the 30°C set
point programmed for Step 3 is reached.
3
30°C
15 minutes Fluid temperature is maintained at 30°C for 15 minutes.
4
30°C
45 minutes
Controller slowly heats fluid until 50°C set point
programmed for Step 5 is reached. Ramp rate is based
on the 45 minute step duration.
5
50°C
1 second
Controller then cools fluid as fast as possible until 5°C set
point programmed for Step 6 is reached.
6
5°C
1 second
“Complete” appears on display. 5°C fluid temperature is
maintained until set point is changed.
0
5
10
15
20
25
30
35
40
45
50
55
0
5
15
30
45
60
75
90
105 120 135 150 165
Minutes
D
e
g
re
e
s
C
Step 1
Step 2
Step 3
Step 4
Step 5
Step 6