Brookfield CAP2000 Viscometer User Manual

Page 9

background image

Brookfield Engineering Labs., Inc.

Page 9

Manual No. M02-313-G0812

I.5 Installation

DO NOT lift the viscometer by the handle or head! LIFT only by the base console or

column!

1) Set the viscometer on a clean level bench surface.

2) Remove shipping foam insert (P/N CAP-122) from the plate area on the CAP Viscometer.

Store the foam insert for future use when shipping or transporting CAP Viscometer.

Foam Insert Used

When Shipping

CAP Viscometer

Figure I-2: Detail of Foam Insert

3) Verify that the viscometer’s power requirements match your power source BEFORE

connecting it to power.

The AC input voltage and frequentcy must be within the appropriate range as shown

on the back of the viscometer head.

Note: The CAP Viscometer must be earth grounded. Use the three (3) wire power

cord! Do not alter!

4) Connect the power cord to the viscometer and to the power supply (source).

5) If using a printer, connect the printer cable to the printer port and printer.

Note: Ensure that both the printer and the CAP 2000+ are off when connecting cables.