beautypg.com

Brookfield CAP2000 Viscometer User Manual

Page 15

background image

Brookfield Engineering Labs., Inc.

Page 15

Manual No. M02-313-G0812

II.4 Temperature Control Setting

Press the TEMP key and the current temperature setting will blink. The default temperature on

start-up is 25.0°C on low temperature models and 50.0°C on high temperature models.

The temperature ranges are:

Low temperature: 5°C (or 15°C below ambient, whichever is higher) to 75°C

High temperature: 50°C to 235°C

Use the number keys to type in the required set point. The temperature can be set in increments

of 0.1°C. You can turn off the temperature control on a high temperature unit by entering 0.

Use the ENTER key to accept the new set point.

Note: Thermal equilibrium of the sample and of the spindle must be considered for best

measurement results. Upon powering up the Viscometer or after changing the

temperature set point, allow sufficient time for the plate to reach the desired tem-

perature. It is recommended to have the spindle in contact with the plate prior to

introducing the sample material to ensure that the spindle is also at the temperature

of test. Brookfield recommends using the solvent trap at all times to enhance the

temperature control of the sample material. After inserting the sample material onto

the plate, lower the spindle and solvent trap and allow sufficient time for thermal

equilibrium prior to starting the test.

II.5 Hold Time Settings

Hold Time sets the time period between when the RUN TIME key is pressed and when the spindle

begins to rotate. This time period is normally used to ensure thermal equilibrium of the sample

and spindle. The hold time range is 0 to 999 seconds.

Press the HOLD TIME key and the current hold time will blink on the default screen. Use the

number keys to type in the required hold time and press the ENTER key.

Note: When the hold time is set to zero, it is not displayed on the default screen. Run time

will be shown on the default screen only when hold time is set to zero.

Note: With a run time of zero, the hold time will not be used

II.6 Run Time

Run Time sets the time period of spindle rotation. The run time range is 0 to 999 second.

Press the RUN TIME key and the current run time will blink on the default screen. Use the

number keys to type in the required run time and press the ENTER key.

A Run Time of zero sets the viscometer to infinite run mode. In this mode, the spindle will

rotate at the set speed for as long as the RUN TIME key is pressed. The display will shown

“MANUAL”. When the RUN TIME key is released, the spindle will stop rotating.