II.2 Cone Spindle Selection and Setting

Raise the viscometer handle to its highest position.

The CAP cones have viscosity ranges as shown in Tables III-1 and III-2. After selecting the ap-

propriate cone for the viscosity range to be utilized, carefully attach the cone to the viscometer

as shown in Figure II-3.

Be sure to insert the spindle completely into the adapter sleeve; align the flat (see Figure I-1)

on the spindle with the thumb screw and then tighten.

When using the solvent trap (Figure II-3), connect it to the cone adapter by sliding it up, passing

the slot by the thumb screw and turning the trap clockwise onto the thumbscrew. Slide the cone

up into the adapter as far as it will go and hand lock it in place with the thumb screw. Tighten

the thumb screw firmly and securely.

Thumb Screw

Solvent Trap

Cone Spindle

Figure II-3: Cone Attachment

Press the SPINDLE key. The display will change to the spindle entry screen. Using the number

keys, type in the required spindle number.

Two digits must be entered for the cone number. For cone 01 through 09, the first number

remains as “0”.

Note: The default cone setting on power-up will be the last cone entry prior to shutting

off the viscometer.

After the correct two (2) digits have been entered, press the ENTER key and the cone will be

accepted for viscometer calculations. The screen will display the following message: