beautypg.com

Brookfield CAP2000 Viscometer User Manual

Page 27

background image

Brookfield Engineering Labs., Inc.

Page 27

Manual No. M02-313-G0812

Notes: a. Lower the handle gently. DO NOT FORCE THE CONE ONTO THE

PLATE.

b. The sample must completely cover the face of the cone and extend beyond the

edge of the cone about 1.0 mm (see Figure III.3).

c. Release the solvent trap placing it onto the plate over the cone so it does not

touch the cone shaft.

The user should ensure that the substances placed under test do not release poisonous,

toxic or flammable gases at the temperatures to which they are subjected to during the

testing.

7. Allow the cone, plate and sample to equilibrate to the temperature control setting.

Note: A minimum of one (1) to three (3) minutes equilibrium time is recommended,

depending upon the sample.

8. Set the Run Time for rotating the cone (Section II.6) and the Hold Time.

9. Put the printer on-line (optional, Section II.7).

10. Press the RUN key and execute the viscosity measurement.

Note: Due to the dynamics of shearing a fluid in the CAP “H” series Viscometers, the temperature

display may indicate a deflection from the equilibrium temperature setting as the cone

begins rotating at high shear rates. The temperature display may indicate the temperature

of the plate and the momentary changes show the cycling of the temperature control at

high temperature. The precision of the viscosity measurement is maintained within the

limits specified in Table III.1.

11. Read the results of the sample test on the printer or write down the test conditions and

viscosity results from the viscometer display.

12. Relocate the solvent trap onto the cone adapter and raise the handle. Caution: HOT

surface!

13. It is recommended to remove the cone for cleaning. However, with care, the cone

can be cleaned in place. Caution: HOT surface!

14. Clean the viscometer plate (refer to Section I.8). Caution: HOT surface!

III.7 Computer Control

The CAP 2000+ Viscometer can be operated remotely under PC control when using the

CAPCALC application software.