Brookfield CAP2000 Viscometer User Manual
Page 20
Brookfield Engineering Labs., Inc.
Page 20
Manual No. M02-313-G0812
Table III-3: Accuracy for LOW TORQUE CAP 2000+ Viscometer
Cone Number
≤ 500 RPM
10-100% FSR
01
± 2%
02
± 2%
03
± 2%
04
± 2%
05
± 2%
06
± 2%
07
± 2%
08
± 2%
09
± 2%
10
± 2%
Table III-4: Accuracy for HIGH TORQUE CAP 2000+ Viscometer
Cone Number
≤ 400 RPM
750 RPM
900 RPM
900 RPM
10-100% FSR
10-100% FSR
≤ 50% FSR
> 50% FSR
01
± 2%
± 2%
± 2%
± 4%
02
± 2%
± 2%
± 2%
± 4%
03
± 2%
± 2%
± 2%
± 4%
04
± 2%
± 3%
± 3%
± 6%
05
± 2%
± 4%
± 4%
± 8%
06
± 2%
± 5%
± 5%
± 10%
07
± 2%
N/A*
N/A*
N/A*
08
± 2%
N/A*
N/A*
N/A*
09
± 2%
N/A*
N/A*
N/A*
10
± 2%
± 2%
± 2%
± 2%
* Use of this cone at this RPM is not recommended.
The specification of temperature accuracy on CAP Viscometers is stated below:
LOW TEMP UNITS:
• In ambient conditions of 15°C to 30°C, tmperature accuracy is +/- 0.5°C.
• In ambient conditions of 5°C to 15°C or 30°C to 40°C, temperature accuracy is +/- 1.0°C.
HIGH TEMP UNITS:
• In ambient conditions of 15°C to 30°C, accuracy is +/- 0.5°C when the temperature set point
on the viscometer is 50°C to ≤100°C.
• In ambient conditions below 15°C or above 30°C, accuracy is +/- 2.0°C when the temperature
set point on the viscometer is 50°C to ≤100°C.
• In ambient conditions of 15°C to 30°C, accuracy is +/- 1.0°C when the temperature set point
on the viscometer is 101°C to 235°C.