Brookfield Engineering Labs., Inc.

Page 20

Manual No. M02-313-G0812

Table III-3: Accuracy for LOW TORQUE CAP 2000+ Viscometer

Cone Number

≤ 500 RPM

10-100% FSR

01

± 2%

02

± 2%

03

± 2%

04

± 2%

05

± 2%

06

± 2%

07

± 2%

08

± 2%

09

± 2%

10

± 2%

Table III-4: Accuracy for HIGH TORQUE CAP 2000+ Viscometer

Cone Number

≤ 400 RPM

750 RPM

900 RPM

900 RPM

10-100% FSR

10-100% FSR

≤ 50% FSR

> 50% FSR

01

± 2%

± 2%

± 2%

± 4%

02

± 2%

± 2%

± 2%

± 4%

03

± 2%

± 2%

± 2%

± 4%

04

± 2%

± 3%

± 3%

± 6%

05

± 2%

± 4%

± 4%

± 8%

06

± 2%

± 5%

± 5%

± 10%

07

± 2%

N/A*

N/A*

N/A*

08

± 2%

N/A*

N/A*

N/A*

09

± 2%

N/A*

N/A*

N/A*

10

± 2%

± 2%

± 2%

± 2%

* Use of this cone at this RPM is not recommended.

The specification of temperature accuracy on CAP Viscometers is stated below:

LOW TEMP UNITS:

• In ambient conditions of 15°C to 30°C, tmperature accuracy is +/- 0.5°C.

• In ambient conditions of 5°C to 15°C or 30°C to 40°C, temperature accuracy is +/- 1.0°C.

HIGH TEMP UNITS:

• In ambient conditions of 15°C to 30°C, accuracy is +/- 0.5°C when the temperature set point

on the viscometer is 50°C to ≤100°C.

• In ambient conditions below 15°C or above 30°C, accuracy is +/- 2.0°C when the temperature

set point on the viscometer is 50°C to ≤100°C.

• In ambient conditions of 15°C to 30°C, accuracy is +/- 1.0°C when the temperature set point

on the viscometer is 101°C to 235°C.