P
eak
D
etection
The peak dots in the meters can be utilized in many ways; as a peak indicator
alerting that the level reaches just prior to the clip, or as a clip indicator;
respectively to the setting. Set the DIP switch #5-3 to -3 dB for use of a peak
meter, or to 0 dB for use of a clip meter. Furthermore 0 dB if the source is
percussive +2, 1 or 0.5 dB can be set to rust the peak detection.
AD C
onverter
C
aubration
The ambient temperature and AC voltage drift while using that may affects
on the conversion accuracy. The AD converter has automated calibration.
Manual calibration is also available to make the conversion optimum.
1. Press WORD LENGTH button for 2 seconds or more to enter SHIFT
operation mode.
2. Press CAL button. While the calibration is carried out, ERROR LED lights
and the outputs are muted.
R
esume
and
C
lear
the
S
tored
D
ata
• Resume
The AD converter retains following data while the power is off , and resumes
when turning on. Marked with * are default values that the AD converter
reloads when Memory Clear is implemented.
Peak Hold: OFF*/2 s/lnfinity
Meter Mode: Normal*/ Zoom
Clock Frequency: 48*/ 44.1/96/88.2 kHz
Clock Source; Internal*/External
Word Length; 24*/20/l6 bit
Out Channel Selection: 1-4*/5-8
• Memory Clear
To return the data back to the default, press the POWER switch while holding
down the [FREQ/CH SEL] and [CLOCK/CAL] button simultaneously.