ERROR LIST

ERROR

NO.

DEFINITION

CHECK/REPLACE

Err 1

Mathematical mistake; division by zero.
Cannot calculate calibration factor

Will show when trying to calibrate with no
calibration weight

Err 2

Weight loaded too low. Cannot calibrate

While in calibration procedure add more weight in
load mode

Err 3

Unstable weight. Cannot calibrate

Check the load cells mechanical surroundings and
see nothing touches them and the cables are well
connected

Err 5

Calibration disabled

Toggle the calibration switch to enable

Err 6

High saturation state (high A/D)

The load cell is not connected correctly. Check
the connection, cables, and mechanical
connection. If the problem persists, replace the
set of load cells

Err 7

Low saturation state (low A/D)

See Err 6