Welch Allyn CP 50 Plus 12-lead resting electrocardiograph - User Manual User Manual

Page 49

background image

Item

Specification

Electrodes

Rigorously tested for conductivity, adhesion, and hypoallergenic qualities, and exceed all

AAMI standards

Power cable

Meets or exceeds Type SJT.

Environmental operating

conditions
Temperature
Relative humidity
Atmospheric air-pressure

limits


+10° C to +40° C (+50° F to +104° F)
15 - 95% noncondensing (30 - 70% for printing)
700 - 1060 hPa

Environmental storage

conditions
Temperature
Relative humidity
Atmospheric air-pressure

limits


-20° C to +50° C (-4° F to +122° F)
15 - 95% noncondensing
700 - 1060 hPa

Protection against electric

shock

Class I, internally powered Type BF

Mode of operation

Continuous

*

If you print at a high gain setting, the waveform or calibration marks might be clipped. This clipping does not

comply with clause 51.103.1 of the IEC/EN 60601-2-51 standard. Use a lower gain setting to observe the full

waveform.

**

Per AAMI EC11:1991/(R)2001 Diagnostic Electrocardiographic Devices, Section 3.1.2.1 Disclosure of cautionary

information/ performance characteristics paragraph c) Accuracy of input signal reproduction, the manufacturer

shall disclose the methods used to establish overall system error and frequency response. Welch Allyn has used

methods A & D, as prescribed in section 3.2.7.2 and 4.2.7.2 of this same standard, to verify overall system error

and frequency response. Because of the sampling characteristics and the asynchronism between sample rate and

signal rate, digital ECG systems such as the CP 50 may produce a noticeable modulating effect from one cycle to

the next, particularly in pediatric recordings. This phenomenon is not physiologic.

*** Disposable electrodes from Welch Allyn shall be used during paient defibrillation.

Specifications are subject to change without notice.

Directions for use

Specifications 45