System problems, The electrocardiograph shuts down during printing – Welch Allyn CP 50 Plus 12-lead resting electrocardiograph - User Manual User Manual
Page 42

System problems
The electrocardiograph won't turn on when it is plugged in
Causes
•
The AC power connection is faulty.
•
An AC fuse is blown.
•
There is no AC power.
Actions
•
Check the AC power source.
•
Check the AC fuses.
The electrocardiograph won't turn on when it is unplugged
Causes
•
The battery is disconnected or incorrectly connected.
•
The battery is low, not charging, depleted, or bad.
Actions
•
Check the battery connections.
•
Recharge the battery.
•
Replace the battery.
The electrocardiograph shuts down during printing
Causes
•
The battery is low or bad.
Actions
•
Recharge the battery.
•
Replace the battery.
The electrocardiograph prints fewer than 10 reports on a full battery
charge
Causes
•
The battery is degraded.
Actions
•
Replace the battery.
38 Troubleshooting
CP 50™ and CP 50 Plus™ 12-lead resting electrocardiograph