Maintenance, Cleaning the equipment – Welch Allyn CP 50 Plus 12-lead resting electrocardiograph - User Manual User Manual

Page 35

background image

Maintenance

Cleaning the equipment

WARNING Keep the electrocardiograph, reusable electrodes, and the
patient cable clean. Patient contact with contaminated equipment can
spread infection.

CAUTION Never let soap or water come into contact with the
electrocardiograph’s internal printer, connectors, or jacks.

CAUTION Never immerse the electrocardiograph or the patient cable in
liquid. Never autoclave or steam clean the electrocardiograph or the patient
cable. Never pour alcohol directly on the electrocardiograph or the patient
cable, and never soak any components in alcohol. If any liquid enters the
electrocardiograph, remove the electrocardiograph from service, and have
it inspected by a qualified service person before using it again.

Acceptable cleaning solutions:

Mild detergent and water, ½ tsp detergent per cup of water

Bleach and water, 1 part bleach (6.00% sodium hypochlorite) with 9 parts water

Isopropyl alcohol and water, 70% by volume

PDI

®

Sani-Cloth

®

Plus wipes (14.85% isopropanol)

CaviWipes™ (17.2% isopropanol)

To clean the equipment (monthly or more often if needed)

1.

Disconnect the power plug from the AC outlet.

2.

Turn off the electrocardiograph. (Press and hold the power button for at least six

seconds until the screen goes blank.)

31