The Search button
gives you access to patient data in the test directory or in
a connected database (CardioPerfect workstation or EMR).
•
The Worklist button (CP 50 Plus only) gives you access to patient data in the
worklist.
5. If a Waiting for 10 (or 20) seconds of quality data message appears, at least 10 or
20 seconds of ECG data have been collected with excessive artifact. Time
requirements in the message may vary, based upon selected print format. Minimize
the artifact, as described under Troubleshooting. Then wait for the test to print. If
necessary, you can override the wait time and print the available data immediately,
but the result might be an incomplete or poor-quality test.
6. After the test prints, select the desired option: Reprint, Save, Retest, or Assign
(Any tests that you save to a USB mass-storage device can be retrieved only from a
CardioPerfect workstation.)
WARNING To avoid the risk of associating reports with the
wrong patients, make sure that each test identifies the
patient. If any report does not identify the patient, write the
patient identification information on the report immediately
following the ECG test.
26 ECG tests
CP 50™ and CP 50 Plus™ 12-lead resting electrocardiograph