Item

Specification

Report formats, external

printer*
Auto ECGs


Average cycles


3x4 @ 25 mm/s, 3x4 @ 50 mm/s, 3x4 + 1R @ 25 mm/s, 3x4 + 3R @ 25 mm/s,
3x4 - 5.0s @ 25 mm/s, 3x4 - 5.0s @ 50 mm/s, 6x2 - 5.0s @ 25 mm/s,
12x1 - 5.0s @ 25 mm/s, No print
3x4 + 3R @ 25 mm/s, 3x4 + 3R @ 50 mm/s, 2x6 + 1R @ 25 mm/s,
2x6 + 1R @ 50 mm/s, 3x4 + 2R @ 25 mm/s, 3x4 + 2R @ 50 mm/s,
6x2 + 2R @ 50 mm/s, No print

ECG storage (in test

directory)

At least 25 ECG tests

Patient storage (in

worklist, CP 50 Plus only)

Up to 50 patients

Frequency range

0.3 to 150 Hz

Digital sampling rate

> 1,000 samples/second/channel

Pacemaker detection

ANSI/AAMI EC11

Power requirement

Universal AC power supply 110-240 V~, 50/60 Hz ~, 1.5 A maximum

AC fuses

Time-lag type, 2.0-amp 250-V rating, Littlefuse 0218002P or equivalent

Rechargeable battery

10.8V, 1.9Ah minimum, 3-cells Li-ion “smart” battery. A fully charged battery prints up to 25

ECGs with one extra copy per test.

Filters
High-performance

baseline
Muscle tremor
AC interference


0.5 Hz
35 Hz
50 Hz or 60 Hz

Safety, EMC, and

regulatory compliance

ANSI/AAMI EC11**
CAN/CSA C22.2 No. 601.1
CAN/CSA C22.2 No. 601.1.1
CAN/CSA C22.2 No. 601.1.2
CAN/CSA C22.2 No. 601.1.4
CAN/CSA C22.2 No. 601.2.25

UL60601-1
IEC/EN 60601-1
IEC/EN 60601-1-1
IEC/EN 60601-1-2
IEC/EN 60601-1-4
IEC/EN 60601-1-6
IEC/EN 60601-2-25 ***
IEC/EN 60601-2-51 (2x6 lead arrangement)

Standard connectivity

Mini USB client and USB host, Ethernet (optional - available with CP 50 Plus only)

Connectivity with

electronic medical records

(CP 50 Plus only)

Through the Welch Allyn CardioPerfect workstation software

44 Specifications