Chapter 7 Troubleshooting
Welch Allyn CP 100 Electrocardiograph
Table 4. System Messages (alphabetical order)
System Message
Problems
Actions
“Insufficient battery power to begin.
Please connect AC power and try again.”
Battery low or bad.
•
Recharge the battery. See
•
Replace battery. See
“Insufficient space available”
Not enough space on memory card
or removable USB storage device.
•
Delete some tests from the card or USB storage
device at a PC.
•
Use a different card or USB storage device.
“Memory card error”
Problem writing to memory card
or removable USB storage device.
•
Verify that the write-protect tab is in the
unprotected position.
•
Reseat the card in its slot.
•
Use a different card or USB storage device.
“Out Of Paper”
•
Printer is out of paper.
•
Printer door is open.
•
Load paper. See
•
Close the printer door. See
“Paper Jam”
Paper was loaded incorrectly.
“Powering down”
Low battery.
Recharge the battery. See
“Problem loading the following settings:
*
*
*
Using default settings.”
Problem loading your settings at startup.
May indicate memory problems.
Contact Technical Support. For phone numbers, see
“Shutdown?”
was pressed while printing a
Rhythm ECG.
To shut down, press
.
To cancel the shutdown, press
.
“Temperature Error”
Printer head temp is too high.
Allow to cool, then try again.