Problem-solving suggestions – Welch Allyn CP 100 12-Lead Resting Electrocardiograph - User Manual User Manual
Page 72

68
Chapter 7 Troubleshooting
Welch Allyn CP 100 Electrocardiograph
Problem-Solving Suggestions
This section includes several tables:
•
Lead quality problems (
•
System failure problems (
•
System messages (
Table 4
on page 70)
If you try these suggestions and still have problems, contact Technical Support. For phone
numbers, see
.
Table 2. Lead Quality Problems
Condition
Causes
Actions
A dot is flashing on the Lead Status
screen.
OR
Lead-off information is displayed on the
screen.
OR
One or more leads prints as a square
wave:
•
Electrode contact may be poor.
•
A lead may be loose.
Note: Square waves may indicate that the
electrocardiograph is inoperative, but they
are more likely to indicate loose leads.
•
Reattach the lead.
•
Replace the electrode.
•
Verify that the electrode area has been properly
prepared: shaved, cleaned with alcohol or
acetone, allowed to dry.
•
Verify that electrodes have been properly stored
and handled.
Wandering baseline (an upward and
downward fluctuation of the waveforms):
•
Electrodes that are dirty, corroded,
loose, or positioned on a bony area.
•
Insufficient or dried electrode gel.
•
Oily skin or body lotions.
•
Rising and falling of chest during rapid
or apprehensive breathing.
•
Clean skin with alcohol or acetone.
•
Reposition or replace electrodes.
•
Help patient relax.
•
If wandering baseline persists, turn the baseline
filter on. See
Muscle tremor interference (random,
irregular voltage superimposed on the
waveforms). May resemble or coincide
with AC interference:
•
Patient is uncomfortable, tense,
nervous.
•
Patient is cold and shivering.
•
Exam bed is too narrow or short to
comfortably support arms and legs.
•
Arm or leg electrode straps are too
tight.
•
Help patient get comfortable.
•
Check all electrode contacts.
•
If interference persists, turn the muscle-tremor
filter on. See
•
If interference still persists, the problem is
probably electrical in nature. See the following
suggestions for reducing AC interference.