Chapter 6 Maintaining the Electrocardiograph

Welch Allyn CP 100 Electrocardiograph

Replacing the AC Fuses

If the green LED on the keyboard does not light up when the electrocardiograph is
connected to AC power, you may need to replace one or both of the AC fuses, as follows.
For the fuse value, see

“Fuses”

on page 73.

1.

Unplug the electrocardiograph from AC power if connected.

2.

Use needle-nosed pliers to remove the fuse case. See

Figure 44

.

3.

Inspect the fuses. If either fuse is dark or has a broken wire, replace the fuse.
See

Figure 45

.

4.

Insert the fuse case. Line it up with the opening; it goes in only one way.

WARNING Failure to unplug could result in electrocution.

Figure 44. Removing AC Fuse Case

Figure 45. AC Fuses Removed