Directions for Use

Chapter 5 Performing ECG Tests

55

Adjusting the ECG Waveforms

To adjust the waveforms after printing an Auto ECG or while printing a rhythm ECG,
press an item’s number key as needed until the desired choice appears.

Item

Effect

Change Leads
(available for Rhythm
ECG only)

Cycles through the lead groups three at a time.

Adjust Gain

Cycles through the gain settings in mm/mV (5, 10, 20, AUTO), changing the waveform size.
(AUTO is available for Auto ECGs only, not rhythm ECGs. AUTO is usually the best setting, but
some waveforms may be easier to read on other settings.) To learn how to change the gain’s
default setting, see

“Reviewing the Miscellaneous ECG Settings”

on page 40.

Adjust Baseline Filter

Toggles between the two baseline filter settings (on or off). This filter reduces “wandering
baseline,” an upward and downward fluctuation of the waveforms. It is preferable, if possible,
to eliminate or reduce wandering baseline by addressing the causes, as explained on

page 68

.

To learn how to change the filter’s default setting, see

“Reviewing the Miscellaneous ECG

Settings”

on page 40. Caution: You cannot perform ST segment analysis on waveforms that

were recorded with the muscle filter turned on. For details, see the caution on

page 52

.

Adjust Muscle Filter

Toggles between the two muscle filter settings (on or off). This filter reduces muscle tremor
interference: random, irregular voltage superimposed on the waveforms. It is preferable,
if possible, to eliminate or reduce muscle tremor by addressing the causes, as explained
on

page 68

. To learn how to change the filter’s default setting, see

“Reviewing the

Miscellaneous ECG Settings”

on page 40. Caution: You cannot perform ST segment analysis on

waveforms that were recorded with the muscle filter turned on. For details, see the caution on

page 52

.

Adjust Speed
(available for Rhythm
ECG only)

Cycles through the paper speed settings in mm/sec (10, 25, 50).

5 mm/mV

20 mm/mV

10 mm/mV

Wandering baseline

Muscle tremor interference

50 mm/sec

25 mm/sec

10 mm/sec