Adjusting the ecg waveforms, Adjust baseline filter, Adjust muscle filter – Welch Allyn CP 100 12-Lead Resting Electrocardiograph - User Manual User Manual
Page 59: Adjust gain, Adjust speed

Directions for Use
Chapter 5 Performing ECG Tests
55
Adjusting the ECG Waveforms
To adjust the waveforms after printing an Auto ECG or while printing a rhythm ECG,
press an item’s number key as needed until the desired choice appears.
Item
Effect
Change Leads
(available for Rhythm
ECG only)
Cycles through the lead groups three at a time.
Adjust Gain
Cycles through the gain settings in mm/mV (5, 10, 20, AUTO), changing the waveform size.
(AUTO is available for Auto ECGs only, not rhythm ECGs. AUTO is usually the best setting, but
some waveforms may be easier to read on other settings.) To learn how to change the gain’s
default setting, see
“Reviewing the Miscellaneous ECG Settings”
Adjust Baseline Filter
Toggles between the two baseline filter settings (on or off). This filter reduces “wandering
baseline,” an upward and downward fluctuation of the waveforms. It is preferable, if possible,
to eliminate or reduce wandering baseline by addressing the causes, as explained on
.
To learn how to change the filter’s default setting, see
“Reviewing the Miscellaneous ECG
on page 40. Caution: You cannot perform ST segment analysis on waveforms that
were recorded with the muscle filter turned on. For details, see the caution on
.
Adjust Muscle Filter
Toggles between the two muscle filter settings (on or off). This filter reduces muscle tremor
interference: random, irregular voltage superimposed on the waveforms. It is preferable,
if possible, to eliminate or reduce muscle tremor by addressing the causes, as explained
on
. To learn how to change the filter’s default setting, see
on page 40. Caution: You cannot perform ST segment analysis on
waveforms that were recorded with the muscle filter turned on. For details, see the caution on
Adjust Speed
(available for Rhythm
ECG only)
Cycles through the paper speed settings in mm/sec (10, 25, 50).
5 mm/mV
20 mm/mV
10 mm/mV
Wandering baseline
Muscle tremor interference
50 mm/sec
25 mm/sec
10 mm/sec