Welch Allyn CP 100 12-Lead Resting Electrocardiograph - User Manual User Manual
Page 38

34
Chapter 4 Reviewing the ECG Settings
Welch Allyn CP 100 Electrocardiograph
Item (in
Figure 21
on page 33)
Description
A. Patient data
See
“Reviewing the Patient Data Fields Available”
B. ECG measurements
Standard.
C. Interpretation (optional)
See
“Reviewing the Interpretation and Copy Settings for Auto Reports”
D. Report status label
“Reviewing the Interpretation and Copy Settings for Auto Reports”
E. 3 rows, 4 columns
“Reviewing the Format Settings for Auto Reports”
F.
Rhythm leads
“Reviewing the Format Settings for Auto Reports”
G. Software version
“Reviewing the Device Information”
H. Device ID
I.
Practice ID
J. Date and time
See
K. AC filter
See
“Mains Filter”
on page 40.
L.
Frequency range
Lower limit: baseline filter on = 0.5, off = 0.3
Upper limit: muscle filter on = 35, off = 150
See
on page 55 and
M Gain
N. Paper speed
O. Calibration pulse
Amplitude reference — represents the current height of a one-millivolt signal.
Is adjusted for the selected gain:
5 mm/mV = 0.5 x
10 mm/mV = 1 x
20 mm/mV = 2 x