Chapter 6 Maintaining the Electrocardiograph

61

Replacing the Battery

If you have recharged the battery and the electrocardiograph still does not turn on when
unplugged, or if the battery loses its charge quickly, replace the battery as follows.
(For part number, see

“Accessories”

on page 6.)

1.

Unplug the electrocardiograph from AC power if connected.

2.

Turn the electrocardiograph upside-down.

3.

Unscrew and remove the battery door. See

Figure 38

.

4.

Lift out the battery. See

Figure 39

.

5.

Inspect the fuse. See

Figure 40

.

If the “Z” wire is intact, go to

Step 6

.

If the “Z” wire is broken or dark, replace the fuse. See

“Replacing the Battery

(DC) Fuse”

on page 63.

Figure 40. Battery Fuse Plugged In

Figure 38. Removing the Battery Door

Figure 39. Lifting Out the Battery

“Z” wire