Wacs hl7 interface option, Outbound messaging – Welch Allyn Connectivity Server - Programmers guide User Manual

Page 62

background image

58

Contacts and specifications

Welch Allyn Connectivity Server

WACS HL7 Interface option

Outbound messaging

Table 27. Configuration variables under HL7 tab

HL7 Manager page

Default setting

Alternate settings

Numeric data filter

Median (can be mean for
even number of samples)

Closest (most recent)

Numeric data filter for NIBP

Closest (most recent)

Median (can be mean for
even number of samples)

HL7 - General Settings page

HL7 versions supported

2.4

2.31, 2.3

Allow observations for multiple patients in one HL7 message

Off

On

Number of seconds to be retrieved when a new patient is
connected

7200

Field entry allowed

Allow multiple observations for a patient in one HL7 message

Off

On

Number of times the server will try to send an HL7 message to
client before disconnecting

5

Field entry allowed

Message delay (seconds) (for push)

420

Field entry allowed

HL7 - Push on Event Settings page

Each vital sign listed in page

Off

On

HL7 - Observation Identifiers page

Vital-sign labels

As listed in page

Field entry allowed

Table 28. Non-supported features

Simultaneous clients: WACS sends data to only one CIS

HL7 batch processing: WACS only sends data at a regular interval or in response to a query

Wildcards: WACS does not accept wildcards