beautypg.com

Unacceptable adt messages and wacs error messages – Welch Allyn Connectivity Server - Programmers guide User Manual

Page 57

background image

Programmer’s Guide

Inbound ADT HL7 messages

53

Unacceptable ADT messages and WACS error messages

WACS returns error messages to the CIS/HIS ADT server in response to these messages:

Message types other than A01, A03, A04, A05, A08, A011 or A13, and your system
has not been programmed to block messages other than these.

HL7 messages containing noncompliant format. In these cases, WACS rejects the
entire content of the message.

WACS error message: non-accepted message type

WACS error message:

MSH|^~\&|||||20090203130308.435-0500||ACK|16249|P|2.4
MSA|AR|613958749|No appropriate destination could be found to which this message could be
routed.
ERR|^^^207&Application Internal Error&HL70357

ADT/ACK error message: noncompliant format

In this example, the telephone number (highlighted below in bold text) is a noncompliant
format.

ADT^A01 message:

MSH|^~\&|AccMgr|1|||20090130160820||ADT^A03|6139110482|P|2.3.1EVN|A03|20090130160819
PID|1|841940^^^AccMgr^PN|500138979^^^AccMgr^MR^1||SAMPLE^PATIENT^L||19920722|F|T
EST TEST|W|2001 SAMPLE
RD^^SAMPLE^NY^132110000^^M|31|8005551212||E|S|NO|6943809^^^AccMgr^VN^1|099809
553|||2|USA||||NOT A VETERAN|||N
PV1|1|O|2E^237^02^1|3||2E^237^02|1370^TEST^TEST^^^^^^AccMgr^^^^CI|1370^ TEST ^
TEST ^^^^^^AccMgr^^^^CI||TWM||||1|||1370^
WESTPFAL^EDITH^^^^^^AccMgr^^^^CI|67|6943809^^^AccMgr^VN^1|200^TEST||||||||||||||||1|1||
1||P|||20081008111700|20081009180000|3083.9|3083.9||||0
PV2||SOB^NO||||||20081008111700

WACS error message:

MSH|^~\&|||||20090130160831.875-0500||ACK|5804|P|2.4
MSA|AE|6139110482|The phone number component must be supplied and should be in the
following format [999-9999].
ERR|^^^207&Application Internal Error&HL70357

Note

The Acuity System accepts middle initial only (not middle name) in messages.