Unacceptable adt messages and wacs error messages – Welch Allyn Connectivity Server - Programmers guide User Manual
Page 57

Programmer’s Guide
Inbound ADT HL7 messages
53
Unacceptable ADT messages and WACS error messages
WACS returns error messages to the CIS/HIS ADT server in response to these messages:
•
Message types other than A01, A03, A04, A05, A08, A011 or A13, and your system
has not been programmed to block messages other than these.
•
HL7 messages containing noncompliant format. In these cases, WACS rejects the
entire content of the message.
WACS error message: non-accepted message type
WACS error message:
MSH|^~\&|||||20090203130308.435-0500||ACK|16249|P|2.4
MSA|AR|613958749|No appropriate destination could be found to which this message could be
routed.
ERR|^^^207&Application Internal Error&HL70357
ADT/ACK error message: noncompliant format
In this example, the telephone number (highlighted below in bold text) is a noncompliant
format.
ADT^A01 message:
MSH|^~\&|AccMgr|1|||20090130160820||ADT^A03|6139110482|P|2.3.1EVN|A03|20090130160819
PID|1|841940^^^AccMgr^PN|500138979^^^AccMgr^MR^1||SAMPLE^PATIENT^L||19920722|F|T
EST TEST|W|2001 SAMPLE
RD^^SAMPLE^NY^132110000^^M|31|8005551212||E|S|NO|6943809^^^AccMgr^VN^1|099809
553|||2|USA||||NOT A VETERAN|||N
PV1|1|O|2E^237^02^1|3||2E^237^02|1370^TEST^TEST^^^^^^AccMgr^^^^CI|1370^ TEST ^
TEST ^^^^^^AccMgr^^^^CI||TWM||||1|||1370^
WESTPFAL^EDITH^^^^^^AccMgr^^^^CI|67|6943809^^^AccMgr^VN^1|200^TEST||||||||||||||||1|1||
1||P|||20081008111700|20081009180000|3083.9|3083.9||||0
PV2||SOB^NO||||||20081008111700
WACS error message:
MSH|^~\&|||||20090130160831.875-0500||ACK|5804|P|2.4
MSA|AE|6139110482|The phone number component must be supplied and should be in the
following format [999-9999].
ERR|^^^207&Application Internal Error&HL70357
Note
The Acuity System accepts middle initial only (not middle name) in messages.