beautypg.com

Data exported from the acuity system to wacs – Welch Allyn Connectivity Server - Programmers guide User Manual

Page 13

background image

Programmer’s Guide

System overview

9

Data exported from the Acuity System to WACS

The Acuity System exports the following vital-sign data:

Table 4. Numeric vital-sign data exported from Acuity Systems to WACS

Vital sign

Units

Default units

Preferred HL7 tag
(OBX-3)

Heart rate

BPM (1/min)

BPM (1/min)

Heart Rate

Temperature 1, 2

DegF (°F),
DegC (°C)

DegC (°C)

Temperature

SpO

2

Per cent (%)

Per cent (%)

SPO2

CO

2

In

mmHg, kPa, %

mmHg

CO2(In)

CO

2

Ex

mmHg, kPa, %

mmHg

CO2(Ex)

RESP

Br/M (1/min)

Br/M (1/min)

RR/BR

NIBP (DIA / SYS / MEAN)

mmHg

mmHg

NIBP

IBP - 1 (DIA / SYS / MEAN)

unlabeled
arterial
intracranial
umbilical artery
umbilical vein
pulmonary artery
central venous

mmHg

mmHg

IBP1
ART
ICP
UA
UV
PA
CVP

IBP - 2 (DIA / SYS / MEAN)

unlabeled
arterial
intracranial
umbilical artery
umbilical vein
pulmonary artery
central venous

mmHg

mmHg

IBP2
ART
ICP
UA
UV
PA
CVP

PVCRate

PVC/Min (1/min)

PVC/Min (1/min)

PVC