Programmer’s Guide

HL7 standard protocols

13

Time/date in HL7 segments

When importing vital-sign data into an electronic patient-charting application, use the
timestamp associated with the OBR for the vital-signs record in the chart.

Format

Date and time in the HL7 segments are represented in this format:
YYYYMMDDHHMMSS.mmm±TZTZ

Displayed time is local.

Hours are in the range 0-23.

Seconds (SS) are followed by the following:

3-digit millisecond value (which appears as 000 if no other value is provided)

+ or a -

4-digit time-zone code

Example of HL7 time/date format:

26 January 2009, at 0.193 seconds past 12:22:02 PM, Pacific time (8 hours behind GMT):
20090126122202.193-0800