beautypg.com

Obx segment example – Welch Allyn Connectivity Server - Programmers guide User Manual

Page 24

background image

20

WACS outbound HL7 protocols

Welch Allyn Connectivity Server

HL7 attributes of OBX (observation/result) segments

OBX segment example

OBX|1|ST|Heart Rate^Heart Rate^WAP||80|^BPM||||||||20090127093400.000-0800

Table 10. HL7 attributes – OBX – observation/result segment

SEQ

LEN

DT

OPT

RP/# TBL#

ITEM#

Element name

WACS
requirement

1

4

SI

O

00569

Set ID - OBX

R

2

2

ID

C

0125

00570

Value Type

R

3

250

CE

R

00571

Observation Identifier

R

4

20

ST

C

00572

Observation Sub-ID

C

5

65536

a

a.

The length of the observation field is variable, depending upon value type. See OBX-2 value type.

*

C

Y

b

b.

May repeat for multipart, single answer results with appropriate data types, e.g., CE, TX, and FT data types.

00573

Observation Value

R

6

250

CE

O

00574

Units

R

7

60

ST

O

00575

References Range

C

8

5

IS

O

Y/5

0078

00576

Abnormal Flags

C

9

5

NM

O

00577

Probability

I

10

2

ID

O

Y

0080

00578

Nature of Abnormal Test

N

11

1

ID

R

0085

00579

Observation Result Status

N

12

26

TS

O

00580

Date Last Observation Normal
Value

N

13

20

ST

O

00581

User Defined Access Checks

N

14

26

TS

O

00582

Date/Time of the Observation

R

15

250

CE

O

00583

Producer's ID

N

16

250

XCN

O

Y

00584

Responsible Observer

N

17

250

CE

O

Y

00936

Observation Method

N

18

22

EI

O

Y

01479

Equipment Instance Identifier

N

19

26

TS

O

01480

Date/Time of the Analysis

N