beautypg.com

Xml interface – Welch Allyn Connectivity Server - Programmers guide User Manual

Page 59

background image

7

55

XML interface

The optional Third-Party Data Stream Interface option provides a means of moving patient
identification and patient numeric data between the Acuity Central Station and a facility’s
information system.

To receive XML messages from WACS, a CIS/HIS opens a TCP/IP socket and listens.
When the Welch Allyn Connectivity Server detects an event, it connects to the socket and
sends a message.

The interface provides the following parameters:

Table 24. XML interface parameters

Parameter

Content

Note

Command

Set
Reset

When the alarm starts.
When the alarm ends.

SourceID

[Patient ID or
Amended ID]

Acuity uses this as the unique patient identifier across
the Acuity network.

Sensitivity

Lethal
High
Parameter

ApplicationID

Welch Allyn

Text

[free-format text]

Alarm type. A valid alarm string (e.g., Heart Rate 250).

MessageID

Event ID

For WACS internal use only.

Timestamp

[time of the alarm]

GuestName

[Full name of the patient]

GuestPhysicalLocation

[Assigned location of the patient]

Unit and room number.

Parameter

[Model name and serial number of
the alarming monitor]

DestinationList

[Target recipient ID]

Present only if:
- Clinician Notifier option is licensed
- Event Delivery Mode = Escalation
- At least one Primary Respondent assigned