Transfer patient names to the screener – Welch Allyn OAE Data Manager - User Manual User Manual

Page 15

background image

6. On the Records section of the Home tab, click New Patient to add patient names

and demographic information for patients tested.

7. On the Records section of the Home tab, click Assign to Patients, and then drag all

unnamed test results on the left side of the window to the appropriate patient name

on the right side of the window.

8. Click OK.

Use Bluetooth to transfer results from the screener to a computer

1. Connect a computer to the screener by pairing the screener with the computer.

The screener displays the message, “Waiting for PC,” detects the connection to the

computer, and waits for an action or communication from the data manager.

2. If a connection is not detected, click Detect Com Port to establish the Com Port

connection.

3. Click the RESULTS icon on the data manager, and then immediately press the UP

control panel button on the screener to display the Bluetooth menu for data transfer.

4. Press the LEFT or RIGHT control panel button on the screener to transfer data.
5. When all data has been transferred, click OK.
6. Navigate to the Recent Results tab, and then click the "Unnamed” patient to view

all transferred tests.

7. On the Records section of the Home tab, click New Patient to add patient names

and demographic information for patients tested.

8. On the Records section of the Home tab, click Assign to Patients, and then drag all

unnamed test results on the left side of the window to the appropriate patient name

on the right side of the window.

9. Click OK.

Transfer patient names to the screener

1. Power on the screener.
2. Connect the screener to a computer using a Bluetooth connection or the micro-USB

connector.
The data manager defaults to the previously used Com Port number and

automatically detects the screener on that port. The message, “Waiting on

computer” appears on the screen. See “Detect Com Port” if this is the first time the

instrument is connected.

3. On the Home tab, click Names.

The Store Names to Instrument window appears.

4. If necessary, click Clear List to clear previously transferred patient names, and then

immediately press the UP control panel button on the screener to display the

Bluetooth menu for data transfer.

5. Press the LEFT or RIGHT control panel button on the screener to transfer data.
6. If desired, click Clear Instrument to delete all patient names and test records on the

screener.
The patient names will be displayed on the device in alphabetical order.

Directions for use

Understand the data manager interface 11