4 set unit, 5 minimum readability setting, 6 date display – Rice Lake TA Series, Rice Lake Weighing Systems User Manual
Page 56: 7 time stamp output, Set unit, Minimum readability setting, Date display, Time stamp output

52 RL-TA Series Operation Manual
16.4 Set Unit
This is a function to allow you to select two units (unit A and unit B) and switch between the units.
Which unit to display can be set in Function 1. You can also set the same unit for unit A and unit B.
Pressing the Function key switches the unit between unit A and unit B.
To use unit B, select the weighing machine function in Function 1 [
1. Set 1]. Unit A can be used in all
the measurement types of the balance.
16.5 Minimum Readability Setting
This function changes the minimum readability. The increment of the minimum readability changes in
the order of 1, 2, and 5. The larger the minimum readability becomes, the less the balance is affected by
disturbance. In addition, it takes less time for the balance reading to become stable.
You can also set the same unit for unit A and unit B in “Section 16.3, Auto Power Off” and switch the
minimum readability between unit A and unit B.
For example, when grams are used the minimum readability is as follows:
Function 1
C2. d.A
Minimum
Readability
C4. d.b
Setting
values
1
0.0001 g
2
0.0002 g
3
0.0005 g
4
0.001 g
5
0.002 g
16.6 Date Display
This is a function to set the date format to be displayed in the balance or output to a printer.
Set date format in Function 1 [
F. dAtE *]:
16.7 Time Stamp Output
This is a function to output measurement data with the current time. You should set the time before this
function is used.
To use this function, enable it in Function 1 [
G. t.o. 1].