Target key
Meanings
Description
Corresponding settings in
Function 1
6. EnU.
5. rE.
FAST
The balance responds quickly but it
is easily affected by vibration
0
1
NORMAL
The balance’s response is
somewhere between fast and slow
0
3
SLOW
The balance responds slowly but is
not easily affected by vibration
1
3
(Refer to “Section 5: Function 1” on page 15 for further information on Function 1.)
In
,
[
0] indicates a fast response. Use this setting when the
balance is subject to a low level of vibration.
[
1] indicates a slow response. Use this setting when the balance
is subject to a high level of vibration.
In [
,
[
1] indicates the fastest response. Use this setting when
the balance is subject to a low level of vibration. As the number
approaches [
5], the balance’s response becomes slower but it is
also less affected by vibration.
* To adjust the response more finely, use Function 1.
This function is unavailable when the gravimeter or addition function is enabled.
4.7 Data Output
Measured data and GLP data can be output to a personal computer, printer, or similar device via the
built-in RS-232C interface. In the factory default setting, pressing the Print key once will output the
current measured value when the measurement becomes stable.
is displayed while the data is being output.
Refer to “Section 5.5: Interface” on page 20 for further information on the output settings.
4.8 Hanging Measurement
The balance is equipped with a hanging hook to suspend a hanging pan for weighing. Perform hanging
measurements when weighing electromagnetic, electrostatic, or high-temperature samples that cannot be
accurately weighed on the normal pan.
Remove the pan and gently tilt the balance backward.
Take care that the door does not fall off.
(Continued on next page.)