7 advice cal and full-automatic span adjustment, 8 auto repeatability measurement (arm), Advice cal and full-automatic span – Rice Lake TA Series, Rice Lake Weighing Systems User Manual
Page 52: Adjustment, Auto repeatability measurement (arm) 48

48 RL-TA Series Operation Manual
☆ Hints ☆
If more than one weight is used in combination, enter the total error of the weights to be used.
The range of the weight error should be within
±100.00 mg. If any value outside the range is entered, it
will result in the display of [
r - Err]. When [r - Err] is displayed, press any key to return the display
to Step 2.
14.7 Advice CAL and Full-automatic Span Adjustment
When Advice CAL determines that a certain amount of time has elapsed or that the temperature has
changed since the previous span adjustment, the “CAL” sign flashes, indicating that the balance needs to
be calibrated. Perform a span adjustment immediately when the “CAL” sign begins flashing.
To use this function, set Function 1 to [
E. AdC 1].
“Full-automatic span adjustment” is available only with the HTR series.
This function automatically performs a span adjustment with built-in weights when “Advice CAL”
determines that the balance needs to be calibrated, causing the “CAL” sign to flash.
The span adjustment using built-in weights will be performed about 10 minutes after the [CAL] sign
begins flashing, once the balance has stabilized with no load.
“Full-automatic span adjustment” can be performed only when [
8. CA. 1] (span adjustment with built-in
weights) is enabled.
14.8 Auto Repeatability Measurement (ARM)
This is a function to calculate a standard deviation by loading and unloading the built-in weight and
repeating span adjustment 10 times. Stable measurement can be performed by adjusting the response
through the single-touch response setting function.
Press and hold the Cal key for a few seconds. When
[
ArN] is displayed, release the key.
After a while, the display switches to [
Cnt. 1] and the
measurement starts. As the repeatability measurement
proceeds, the number of measurements displayed in
[
Cnt. *] increases accordingly until it reaches [10].
When the measurement is completed, the standard deviation, for example, [
0.000073g], is displayed.
Pressing any key returns the display to measurement mode.
To cancel the measurement, press the Print key.
Refer to page 12 for setting the single-touch response.