10 unit converting, Displaying the coefficient setting, Enter a coefficient – Rice Lake TA Series, Rice Lake Weighing Systems User Manual
Page 29: Changing numbers, Select digit places to specify a coefficient, Save the coefficient, Put a sample to weigh on the balance, Unit converting

RL-TA Series Operation Manual 25
10 Unit Converting
The function to convert units multiplies the weight of a sample on the balance by the saved coefficient, and
displays the result.
* Select [
1 Set . 4] (Unit Converting) in Function 1.
1.
Displaying the coefficient setting.
Make sure that the [#] sign is displayed and press and hold the
Function key for a few seconds.
(This operation may be performed with a container placed on
the balance.)
The display then blinks [
C. SEt].
The previously-saved reference sample weight blinks.
2.
Enter a coefficient.
Press the Zero/Tare key.
The digit furthest to the right blinks.
3.
Changing numbers.
Pressing the Zero/Tare key changes the number as
shown on the right.
4.
Select digit places to specify a coefficient.
Pressing the Function key shifts the blinking digit to the left,
allowing you to set a higher-order digit.
After a number is set, the minus sign [M] blinks. Pressing the
key again makes the digit furthest to the right blink.
5.
Save the coefficient.
Press the Target key.*
(* Pressing the Print key cancels the setting
of a reference sample weight.)
6.
Put a sample to weigh on the balance.
Put a sample to weigh on the balance.
The weight of the sample is multiplied by the coefficient, and
the result is displayed.
(Continued on next page.)