0 calibration, 1 front panel calibration, 2 edp command calibration – Rice Lake SURVIVOR 420HE Hostile Environment - Installation Manual User Manual
Page 31: Section 4.1 on

420HE Installation Manual - Calibration
27
4.0
Calibration
The
420HE
can be calibrated using the front panel, EDP commands, or the
Revolution
®
configuration utility. Each
method consists of the following steps:
•
Zero calibration
•
Entering the test weight value
•
Span calibration
•
Optional rezero calibration for test weights using hooks or chains.
The following sections describe the calibration procedure for each of the calibration methods.
XXXXXXX
XXXXXXX
XXXXXXX
XXXXXXX
WZERO
WV AL
WSPAN
ALGOUT
DIGI N
SETPNT
PROGRM PFORMT
SERIAL
CALIBR
CONFIG FORMAT
REZERO
VERS
Figure 4-1. Calibration (CALIBR) Menu
4.1
Front Panel Calibration
To calibrate the indicator using the front panel, do the 
following: 
1. Place the indicator in setup mode (display
reads
CONFIG
) and remove all weight from the
scale platform. If your test weights require 
hooks or chains, place the hooks or chains on 
the scale for zero calibration. 
2. Press
until the display reads
CALIBR
(see
Figure 4-1). Press to go to zero calibration 
(
WZERO
).
3. With
WZERO
displayed, press to calibrate
zero. The indicator displays
*CAL*
while
calibration is in progress. When complete, the 
A/D count for the zero calibration is displayed. 
DO NOT adjust this value after WSPAN has 
been set! Press again to save the zero 
calibration value and go to the next prompt 
(
WVAL
).
4. With
WVAL
displayed, press to show the
stored calibration weight. Use the procedure 
shown in Figure 3-4 15 to enter the actual 
value of the test weights to be used for the span 
calibration. Press to store the entered 
WVAL
value and go to span calibration (
WSPAN
).
5. With
WSPAN
displayed, place test weights on
the scale and press to calibrate span. The 
indicator displays 
*CAL*
while calibration is in
progress. When complete, the A/D count for 
the span clibration is displayed. Press again 
to store the span calibration value and go to the 
next prompt (
REZERO
).
6. The rezero function is used to remove a
calibration offset when hooks or chains are 
used to hang the test weights. 
• If no other apparatus was used to hang the test
weights during calibration, remove the test 
weights and press to return to the CALIBR 
menu. 
• If hooks or chains were used during
calibration, remove these and the test weights 
from the scale. With all weight removed, press 
to rezero the scale. This function adjusts the
zero and span calibration values. The indicator 
d i s p l a y s 
* C A L *
w h i l e t h e z e r o a n d sp a n
calibrations are adjusted. When complete, the 
adjusted A/D count for the zero calibration is 
displayed. Press to save the value, then press 
to return to the CALIBR menu.
7. Press
until the display reads
CONFIG
, then
press to exit setup mode.
4.2
EDP Command Calibration
To calibrate the indicator using EDP commands, the 
indicator EDP port must be connected to a terminal or 
personal computer. See Section 2.3.3 on page 8 for 
EDP port pin assignments; see Section 5.0 on page 29 
for more information about using EDP commands.
Once the indicator is connected to the sending device, 
do the following:
1. Place the indicator in setup mode (display
reads
CONFIG
) and remove all weight from the
scale platform. If your test weights require 
hooks or chains, place the hooks or chains on 
the scale for zero calibration. 
2. Send the WZERO EDP command to calibrate
zero. The indicator displays
*CAL*
while
calibration is in progress.