1 service counters, Service counters – Rice Lake MSI-3460 Challenger 3 User Manual
Page 54

50 MSI-3460 Operator’s Manual
8.1.1
Service Counters
The MSI-3460 maintains two service counters for safety.
• The first counter counts the number of times the scale has been overloaded.
• The second counter counts lifts above 25% of capacity.
These counters serve to warn the user to inspect the load train after a number of
overloads, also when there is a chance of fatigue failure. The power up routine will be
interrupted when the lift counter exceeds 16383 lifts or the overload counter exceeds
1023 overloads. If the screen displays LFCNT when unit is powered on :
1. Push TARE to display the 25% lift counter.
2. Push TARE again to see the overload lift counter.
3. Push the ZERO key to acknowledge the warning and return to standard scale
operation.
WARNING
Note
The power up warning message won't appear again for another 16383
lifts ( or 1023 overloads).
To access the service counters:
4. Program a user function key to be TEST (see function key setup).
5. Press TEST (USER).
6. Within two seconds of pressing the TEST (USER) key, press TARE.
7. The display flashes “LFCnt” (for Lift Counter) followed by the number of times
the weight has exceeded 25% of capacity.
8. Press TEST (USER). The display flashes “OLCnt” (for Overload Counter)
followed by the number of times the weight has exceeded capacity.
9. Press TEST (USER). The display flashes the C-Cal value.
10.Press TEST (USER). Returns to standard weighing mode.
Only a MSI factory representative can reset the service counters,
as these are important safety warning features. Depending on
the circumstances, a thorough load train inspection might be
necessary to ensure user safety.
Reference MSI’s “Crane Scale Safety and Periodic Maintenance manual” (Pub.
243-08-94D) for proper loading techniques to improve the safety and longevity of
your
MSI-3460
crane scale. This publication is available at www.msiscales.com
and is included in the CD shipped with your crane scale.
Auto total does not
work
Weight must be stable
Wait for the motion light to go
out, or increase filtering for
more stability
Weight thresholds not
reached
You must exceed 1% of
capacity for autototal to work.
Then you must drop below
0.5% of capacity for
additional weighments to
register.
Problem
Possible Cause
Solution
Figure 8-1. Troubleshooting