Msi-9300 porta-weigh+ crane scale – Rice Lake MSI9300 PortaWeigh Crane Scale Operator Manual User Manual
Page 28

Page 28 MSI-9300 Crane Scale™
•
User Guide
MSI-9300 Porta-Weigh+ Crane Scale
LEGAL-FOR-TRADE CONTROL DEVICE
When the MSI-9300 is configured for NTEP or OIML Legal-For-Trade (LFT) only one remote device is
allowed to have control over the CellScale. The control device is the only RF connected device that can
zero, tare, calibrate, and any other controls that are considered metrologically significant. This control can be
assigned to a 3750CS Remote Display, a 9750A Handheld Display, or to a 9020 CellModem connected to a
computer using the MSI Host commands. 9850 Indicator approval is pending.
Additional RF connected devices are allowed, but they can only function as displays with no controls. Thus
you can assign, for example, a 9750A as the control device, but you can still operate a scoreboard display con-
nected to a CellModem (MSI-9020). Additional 9750A’s can monitor the weight, but they would not be able to
zero or tare the weight, or change the weight mode.
The assigned Control Device RF ID must be known before entering this menu. The number needed is refered
to as “My ID” in a 3750CS, or “My address” in the 9750A. These can be read from the Modem settings in
each device. Of course, the main Network numbers much match between the MSI-9300 and the assigned LFT
control device.
Main CellScale Menu ➔ (2) Communications➔ (9) Assigned Legal-For-Trade Control Device
***** Assign Device for Access Right in legal for trade mode *****
Type in Device ID for Access Right range (0=Host to 254):
(ESC) Exit
1) From the main CellScale Menu, press [2] Communications, then [9] Assigned Legal-For-Trade Control
Device.
2) Use the numeric keys on your terminal to enter the address of the control device. Legal addresses range
from 04 to 251. Use only the “My ID” or “My Address” portion of the designated control device.
3) Press
ENTER
to store the control device address. Press
ESC
to abort the reassignment of the control
device.
4) Confirm the correct control address has been assigned.
5) Press ESC one or more times to return to other menus.
This menu does not effect non LFT standard settings – Industrial, or Metric.