OHMS?

Returns the raw Ohm value from the RTD.

For example:

If when measuring a P100-385 at 0 degrees cel sending
OHMS? would return 1.000000E+02, OHM .

PRES_UNIT

Used to set the pressure unit for the indicated display

For example:

To set the pressure unit to psi on the lower display send
PRES_UNIT LOWER, PSI.

PRES_UNIT?

Returns the pressure unit used when measuring pressure for
each of the 3 displays.

REMOTE

Puts the calibrator in remote mode. From the remote mode the
user can still use the keypad to get back to local unless the
command LOCKOUT was entered before REMOTE. Than the
keypad is totally locked out, and the user has to send the LOCAL
command to get back to local operation.

RTD_TYPE

Sets the RTD type. The following is a list of RTD types the way
they should be entered after the command:

PT385_100;

PT392_100;

PTJIS_100;

CUSTOM;

For Example:

RTD_TYPE PT385_100 sets RTD type to PT100-385

RTD_TYPE?

Returns the RTD type.

For Example:

If the RTD type is PT385_100, RTD_TYPE? Will return
PT100_385.

ST_START

Starts a switch test.

37