Measuring pressure – Rice Lake Condec 3030 User Manual
Page 14
Use the DONE option to save the changes and return home. When
a display is deactivated its configuration is retained. When the
display is activated its configuration is checked against the
configurations of the other currently active displays, if the
configurations are in conflict the recalled display's configuration is
modified to avoid the conflict. If all three displays are deactivated the
LOWER display will come on automatically
2.4.6 Setting the RTD probe type
Use the PROBE TYPE option of the RTD Probe Type main menu to
access the RTD Probe selection menu.
There are four probe types to select from P100-385, P100-392, P100-
JIS and CUSTOM. Use the SELECT option to select the desired
probe type and the DONE option to store the change and return
home.
Note: The default probe type is PT100-385.
2.4.7 Damping
Damping can be turned ON or OFF using the Damping menu
selection. When damping is ON, the calibrator displays an running
average reading of ten measurements. The calibrator makes
approximately 3 readings per second.
3. Measuring Pressure
To measure pressure, connect the calibrator using an appropriate
fitting. Choose a pressure setting for the display being used. The
calibrator is equipped with two internal sensors and many optional
external sensors (EPMs) are available. Be sure to choose the sensor
based on working pressures and accuracy.
Note:
Pressure sensors may be damaged and/or personnel injury may
occur due to improper application of pressure. Please refer to the
table of ranges and resolutions at the back of this manual for
information on overpressure and burst pressure ratings. Vacuum
should not be applied to any gauge pressure sensor. The calibrator
12