12.5 Remote Commands and Error Codes

The following tables list all commands, and their descriptions, that
are accepted by the calibrator.

Table 5: Common Commands

Command Description

*CLS

(Clear status.) Clears the error queue.

*IDN?

Identification query. Returns the manufacturer, model num-
ber, and firmware revision level of the Calibrator.

*RST

Resets the calibrator to the power up state.

Table 6: Calibrator Commands

Command Description

CPRT_COEFA

Sets the custom RTD coefficient A

CPRT_COEFA?

Returns the custom RTD coefficient A

CPRT_COEFB

Sets the custom RTD coefficient B

CPRT_COEFB?

Returns the custom RTD coefficient B

CPRT_COEFC

Sets the custom RTD coefficient C

CPRT_COEFC?

Returns the custom RTD coefficient C

CPRT_R0

Sets the custom RTD R0 resistance

CPRT_R0?

Returns the custom RTD R0 resistance

DAMP

Turns Damp on or off.

DAMP?

Returns if DAMP is on/off

DISPLAY

Turns on/off the displays specified in the command

DISPLAY?

Returns which displays are on/off

FAULT?

Returns the most recent error code

ERROR_LOOP

Turns loop power on or off in percent error mode

ERROR_LOOP?

Returns the current state of loop power in error mode

ERROR_MODE

Turns percent error mode on or off

ERROR_MODE?

Returns whether percent error mode is on or off

ERROR_PORT

Set the pressure port for percent error mode

ERROR_PORT?

Returns the pressure port for percent error mode

FUNC

Sets the display mode as specified in the command

FUNC?

Returns the current mode of the upper, middle, and lower
display

HI_ERR

Sets the 100% of span limit for percent error mode

HI_ERR?

Returns the 100% of span limit for percent error mode

LOCAL

Returns user to manual operation of the calibrator