Serial Commands
73
9.1.2
Reporting Commands
Reporting commands send speciÞc information to the
serial port. The commands listed in Table 9-2 can be
used in either setup mode and normal mode.
9.1.3
Clear and Reset Commands
The following commands can be used to clear and
reset the
920i
:
PCLR
: Program clear. Erases the loaded user program
(setup mode only).
RS
: Reset system. Resets the indicator without
resetting the conÞguration.
RESETCONFIGURATION
: Restores all conÞguration
parameters to their default values (setup mode only).
The
RESETCONFIGURATION
function can also be
initiated by pressing the
Reset Config
softkey under
the VERSION menu.
NOTE:
All load cell calibration
settings are lost when the RESETCONFIGURATION
command is run.
9.1.4
Parameter Setting Commands
Parameter setting commands allow you to display or
change the current value for a particular conÞguration
parameter (Tables 9-3 through 9-11).
Current conÞguration parameter settings can be
displayed in either setup mode or normal mode using
the following syntax:
command
Most parameter values can be changed in setup mode
only; setpoint parameters listed in Table 9-6 on
page 76 can be changed when in normal weighing
mode.
Use the following command syntax when changing
parameter values:
command=value
, where
value
is either a number or a parameter value. Use no spaces
before or after the equal (=) sign. If you type an
incorrect command, the display reads
??
.
For example, to set the motion band parameter on
Scale #1 to 5 divisions, type the following:
SC.MOTBAND#1=5D
For parameters with selectable values, enter the
command and equal sign followed by a question
mark:
command=?
to see a list of those values.
The indicator must be in setup mode to use this
function.
NOTE:
Some parameters are valid only if other
parameters or parameter values are specified. See the
configuration menus in Section
information about parameter dependencies. Restrictions
for front-panel configuration also apply to serial command
configuration.
Command
Function
DUMPALL
List all parameter values
SPDUMP
Print setpoint configuration
VERSION
Write
920i
software version
S#n
Write one frame of stream format to scale
channel n
HARDWARE
Lists option cards installed in slots 1–14.
See Section 10.1.2 on page 88 for more
information about using the HARDWARE
command.
XE
Returns a10-digit code representing any
error conditions currently shown on the
front panel. See Section 10.1.4 on page 89
for more information.
Table 9-2. Reporting Commands
Command
Description
Values
SC.GRADS#n
Graduations
1–9999999
SC.ZTRKBND#n
Zero track band
OFF, 0.5D, 1D, 3D
SC.ZRANGE#n
Zero range
1.9%, 100%
SC.MOTBAND#n
Motion band
1D, 2D, 3D, 5D, 10D, 20D, OFF
SC.SSTIME#n
Standstill time
1-65535
SC.OVRLOAD#n
Overload
FS+2%, FS+1D, FS+9D, FS
SC.DIGFLTR1#n
SC.DIGFLTR2#n
SC.DIGFLTR3#n
Digital filtering
1, 2, 4, 8, 16, 32, 64, 128, 256
SC.DFSENS#n
Digital filter cutout sensitivity
2OUT, 4OUT, 8OT, 16OUT, 32OUT, 64OUT, 128OUT
SC.DFTHRH#n
Digital filter cutout threshold
NONE, 2D, 5D, 10D, 20D, 50D, 100D, 200D, 250D
SC.RATLTRAP#n
Rattletrap filtering
OFF, ON
Table 9-3. SCALES Serial Commands