To transfer data via USB, insert a USB thumb drive into the USB slot. The

thumb drive must be formatted in Base 16 format or it may not work. Base

16 is the default formatting within Windows

®

7. When transferring data

from the EDG, you have your choice of Exporting Job Sites, Exporting

Soil Models or Exporting All. You make your choice by pressing the

corresponding button on the Communication screen.

You can also use the USB device and port to import Soil Models from other

gauges or previous Projects. This is simply accomplished by inserting the

USB device containing desired Soil Models into the port and pressing the

Import Soil Model Button.

The USB port is also used in Updating Firmware for the EDG. This is

accomplished by downloading a Firmware Update from:
http://www.humboldtmfg.com/EDG/Firmware Update.
Place the downloaded file on the USB device and insert it into the USB port

on the gauge. Navigate to the Communication screen and Press the Update

button

under the Heading: Perform Firmware Update.

29. Calibration Check

To run the Calibration Check using the Calibration Check Box. Locate

the Calibration Check Box, which is included with your gauge.

Select the SETUP key on the keypad to open the setup menu. Scroll

thru the menu until you come to “Calibration Check”. Select Calibration

Check, you will see this screen. Follow the screen prompts. First, slide

the Calibration Check Box onto the Soil Sensor and connect them to the

Gauge's Soil Sensor Port.