Bio-Rad SmartSpec Plus Spectrophotometer User Manual
Page 52

Section 8
Troubleshooting
Troubleshooting Table
Problem
Cause
Solution
Unit does not power up
Power switch in off position. 1. Check power switch on back
of unit.
No power from outlet.
2. Check power source.
Bad power cord.
3. Replace.
LCD display difficult
Back lighting burned out.
1. Call Bio-Rad service.
to read
Contrast setting out of
adjustment.
Printer jam
Thermal paper caught
1. Remove excess paper
between print head and
between printer head and plate
thermal plate when tearing
using fine tip tweezers.
off paper.
2. Power down unit, remove
thermal paper roll and re-load
paper in printer.
Printer paper would not
Printer paper inserted
1. Cut a straight clean leading
self feed or advance
incorrectly, leading edge
edge on paper with a pair of
of paper uneven or folded
scissors.
over.
Paper feed button keyboard
2. Call Bio-Rad service.
is faulty.
Print out text too dark
Printer paper damaged
1. Replace with new paper roll.
caused from excessive
heat, moisture.
Paper loaded incorrectly.
2. Be sure the paper is installed
in the proper orientation.
(refer to illustration in Section
9.3)
Thermal print head/plate
3. Replace printer. Call Bio-Rad
damaged.
service.
Print out text too light
1. Paper loaded incorrectly
1. Be sure the paper is installed
or not visible
(incorrect orientation).
in the proper orientation
(refer to illustration in Section
9.3)
2. Printer paper damaged by 2. Replace the printer paper roll.
excessive heat, moisture.
3. Thermal print head/plate
3. Call Bio-Rad Technical
damaged.
Service.
Keyboard buttons not
Incomplete system
1. Power down unit/Restart unit.
responding
initialization during power up.
Pressed wrong button for
operation.
49